RecallDepth

Flexi-Seal Fecal Management System Advanced Odor Control Kit. Rx only; 1 silicone catheter tube assembly, 3-1 Liter collection bags with filters, 1 luer-lock syringe, SKU #411104 ConvaTec Skillman, NJ Made in USA with imported components "For the fecal management of patients with little or no bowel control and liquid or semi-liquid stool."

This recall has been terminated (originally issued December 30, 2008).

Company
Convatec
Recall Initiated
November 20, 2008
Posted
December 30, 2008
Terminated
March 9, 2009
Recall Number
Z-0618-2009
Quantity
795
Firm Location
Skillman, NJ
Official Source
View on FDA website ↗

Reason for Recall

It was discovered that a portion of the inner nylon sleeve of the catheter was visible at the balloon end of a sample device.

Distribution

Worldwide Distribution --- including USA and country of Canada.

Lot / Code Info

Lot number 08-TM-51

Root Cause

Nonconforming Material/Component

Action Taken

Urgent Product Correction Letters were sent by first class mail on November 18, 2008 to Healthcare Professionals and ConvaTec Reps. The reason for the recall was identified along with possible risk to patient. Healthcare professionals were asked to use sufficient lubricating jelly during the insertion process with special care given to those who have a tendency to bleed or are on anti-coagulant medication. Physician should be notified if rectal bleeding or abdominal symptoms develop. ConvaTec reps were asked to contact all health care professionals to whom product or samples were given and reinforce the importance of proper lubrication and following the directions. A return form is to be completed by ConvaTec reps for each visit indicating the name of the Health Care Provider and that the they personally received notice of the product correction. All remaining sample products are to be returned to Besta, Franklin, WI.

More recalls by Convatec

View all recalls by this company →