Unilect ECG Electrodes Unomedical, McAllen Texas. Intended use:Cardiac monitoring

This recall has been terminated (originally issued April 22, 2010).

Company
Convatec
Recall Initiated
March 11, 2010
Posted
April 22, 2010
Terminated
June 21, 2011
Recall Number
Z-1431-2010
Quantity
567,000 units
Firm Location
Skillman, NJ
Official Source
View on FDA website ↗

Reason for Recall

In an emergency situation where defibrillation is necessary and electrodes are already in use, the electrodes may malfunction and incorrectly report lack of cardiac activity.

Distribution

Nationwide distribution: CA

Lot / Code Info

Model Number 4560M; Lots: 706268, 708152, 709705, 711967, 713648, 713977, 714391, 714392, 714685, 714687.

Root Cause

Process design

Action Taken

The one distributor was notified personally on March 9, 2010. All remaining stock was put on hold, and a "URGENT : MEDICAL DEVICE RECALL" letter dated March 11, 2010, was sent to customers. The letter describes the product, problem and action to be taken by customers. All customers were asked to immediately stop distributing and quarantine all recalled lots. The customers were ask to perform a count of recalled product currently in inventory, complete and return the Recall Response Form even if no recalled product is in inventory, send a copy of response to distributor in order to receive reimbursement, and to call 1-800-582-6514 for instructions on how to return the recalled product. If you have any questions, please contact Convatec at 908-904-2425.