RecallDepth

ARTISTE MV Systems (Part Number 8139789), equipped with a syngo RT Therapist (Part Number 8162815) or RT Therapist Assist system (Part Number 8162807), with Software version NET VA10A for RTT4.1.122 or RTT 4.1.125, manufactured by Siemens Medical Solutions, Concord, CA. Medical charged-particle radiation therapy system.

This recall has been terminated (originally issued September 24, 2009).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
April 9, 2009
Posted
September 24, 2009
Terminated
April 12, 2011
Recall Number
Z-1888-2009
Quantity
53 units
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

Unexpected treatment plan alteration-- Software issue may result in unintended modifications to treatment

Distribution

Worldwide Distribution -- US (MA, UT, OH, NJ, NE, FL, OH, MI and PA), Germany, Poland, the UK, Ireland, Italy, France, South Korea, and Belgium.

Lot / Code Info

All Product with Part Number 8139789 (RT Therapists are part of the system).

Root Cause

Software design

Action Taken

Siemens Medical Solutions USA, Inc. notified consignees of the affected product by advisory letter via Certified Mail on April 09, 2009. For further information, contact Siemens Medical Solutions at 1-925-602-8157.

More recalls by Siemens Medical Solutions Usa, Inc

View all recalls by this company →