RecallDepth

Boston Scientific Stonetome Stone Removal Device. Triple Lumen Sphincterotome, 5.5 Fr, 30 mm, 11.5 mm, sterile, Boston Scientific Corp., Natick, MA; REF 3513. The device is used in diagnostic or therapeutic endoscopic retrograde cholangiopancreatography, transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi, to remove stones from the biliary system, and to facilitate injection of contrast medium while occluding the duct with the balloon.

This recall has been terminated (originally issued March 11, 2009).

Company
Boston Scientific Corp
Recall Initiated
January 8, 2009
Posted
March 11, 2009
Terminated
October 22, 2009
Recall Number
Z-1011-2009
Quantity
294 of all products.
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

The ultraviolet light barrier on the pouch may be discolored, which will allow ultraviolet light into the package and may cause the latex retrieval balloon to become brittle and to leak or burst when inflated.

Distribution

Nationwide.

Lot / Code Info

Lot Numbers: 12001574 and 12002381.

Root Cause

Packaging

Action Taken

Consignees were notified via letter dated January 8, 2009 to discontinue use of the affected product and to return it to Boston Scientific. Customers were instructed to complete and fax the Reply Verification Tracking Form to Boston Scientific at 508-683-5578. Direct questions about this recall to your local Sales Representative.

More recalls by Boston Scientific Corp

View all recalls by this company →