Advisor Vital Signs Monitor (model 9200), catalog number 925654335, with Impedance Respiration board (optional circuit board, P/N: 80004B1) installed manufactured with Rev 10 Main boards (P/N: 80002B1) CE/MDD marking, 3-lead 60 Hz ECG, standard SPO2 option, Invasive Pressure/Temperature option, printer and battery.

Company
Smiths Medical Pm, Inc.
Recall Initiated
September 1, 2005
Posted
November 30, 2005
Terminated
June 27, 2007
Recall Number
Z-0233-06
Quantity
5
Firm Location
Waukesha, WI

Reason for Recall

Potential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the primary lead is set to Lead II, III or any of the augmented leads, AVL, AVR or AVF on the Advisor monitor.

Distribution

AZ, CA, KS, KY, NY, TN, TX OUS to countries such as: Egypt, France, Iran, Mexico, Peru, Romania, Saudi Arabia and Spain

Lot / Code Info

serial numbers: AM05070171, AM05070172, AM05070173, AM05070174 and AM05070175

Root Cause

Other

Action Taken

Distirbutors and consignees were notified beginning 09/01/05 via email or phone about Safety Action Bulletin 05-004. Customers were asked to respond to Smiths Medical with a written acknowledgement that they received this Safety Action Bulletin (SAB). The SAB described the issue, actions they should take and instruction on how to have the monitor repaired.