BCI Autocorr Plus Pulse Oximeter/ECG/Respiration Monitor, Catalog Numbers, 3404001 (including S) and 3404002 (including E,F, G, and S) Smiths Medical- a part of Smiths Group plc
- Company
- Smiths Medical Pm, Inc.
- Recall Initiated
- July 18, 2007
- Posted
- September 21, 2007
- Terminated
- October 31, 2008
- Recall Number
- Z-1240-2007
- Quantity
- 178
- Firm Location
- Waukesha, WI
Reason for Recall
Phantom pacer spikes caused by a damaged resistor.
Distribution
World wide: USA including states of KY, MS, PA, LA, and PR and countries of Poland, Mexico, Kuwait, Taiwan, Singapore, Hon Kong, India, Bolivia, Colombia, Chile, Argentia, Brunei Darssalam, Israel, Greece, Romania, Germany, The Netherlands, France, Great Britain, Spain, Brazil, Russia, China.
Lot / Code Info
Catalog number/ serial number: 3404001/AI05110052/ AI05120000/ AI05120011/ AI05120012/ AI05120013/ AI05120018/ AI06060015/ AI06070047/ AI06100004/ AI06100005/ AI06100006/ , 3404001S/AI05110000/ AI05110001/ AI05110002/ AI05110003/ AI05110004/ AI05110005/ AI05110006/ AI05110007/ AI05110008/ AI05110009/ AI05110010/ AI05110011/ AI05110012/ AI05110013/ AI05110014/ AI05110015/ AI05110016/ AI05110017/ AI05110018/ AI05110019/ AI05110020/ AI05110021/ AI05110022/ AI05110023/ AI05110024/ AI05110025/ AI05110026/ AI05110027/ AI05110028/ AI05110029/ AI05110030/ AI05110031/ AI05110032/ AI05110033/ AI05110034/ AI05110035/ AI05110036/ AI05110037/ AI05110038/ AI05110039/ AI05120010/ AI05120016/ AI05120017/ AI05120019/ AI06010042/ AI06010043/ AI06060003/ AI06060004/ AI06060005/ AI06060006/ AI06060007/ AI06060008/ AI06060009/ AI06060010/ AI06060011/ , 3404002/AI05090021/ AI05090024/ AI05110049/ AI05120015/ AI06010035/ AI06010036/ AI06010049/ AI06040000/ AI06040001/ AI06040002/ AI06040003/ AI06040004/ AI06040005/ AI06040006/ AI06040007/ AI06040008/ AI06040009/ AI06040010/ AI06040011/ AI06040012/ AI06040013/ AI06040014/ AI06040015/ AI06040016/ AI06040017/ AI06040018/ AI06040020/ AI06060012/ AI06060013/ AI06060014/ AI06060017/ AI06060018/ AI06060019/ AI06060020/ AI06070037/ AI06070038/ AI06070039/ AI06070040/ AI06070042/ AI06070043/ AI06070044/ AI06070045/ AI06090053/ AI06100011/ AI06100012/ AI07010057/ AI07010058/ , 3404002E/AI05110040 AI05110041/ AI05110042/ AI05120001/ AI05120002/ AI05120005/ AI05120006/ AI05120007/ AI05120008/ AI05120009/ AI05120024/ AI06010041/ AI06010044/ AI06010045/ AI06010046/ AI06010048/ AI06040019/ AI06040023/ AI06040024/ AI06070041/ AI06070046/ AI06070048/ AI06070049/ AI06090050/ AI06090051/ , 3404002F/AI07010067/ AI07010068/ , 3404002G/AI05110050/ AI05110051/ AI05120003/ AI05120004/ AI06060021/ AI06060022/ AI06060023/ AI06070035/ AI06070036/ AI06090054/ AI06090055/ AI06090056/ AI06090057/ AI06090058/ AI06100008/ AI06100009/ AI06100010/ AI06100018/ , 3404002S/AI05110043/ AI05110044/ AI05110045/ AI06010037/ AI06010038/ AI06010039/ AI06010040/ AI06060000/ AI06060001/ AI06060002/ AI06060016/ AI06090052/ AI06100019/ AI06100020/ AI06100021/ AI06100022/ AI06100023/ AI06100024/ AI07010050/ AI07010051/
Root Cause
Other
Action Taken
Customers will be notified via a Customer Information Bulletin CIB Tracking Number: 07-CIB 12, July 18,2007 requesting them to evalute the functionality of their monitor (s). If the monitor does not exhibit any false pace detect makers when pace detect is enabled and respiration is disabled, no further action is required. If it does exhibit false pace detect they will be instructed to contact the Smiths Medical PM, Inc service department. At that time Technical Worksheet, 07-TW02, will be provided. The consignee will need to verify they are capable of perfoming the Technical Worksheet provided. The Smiths Medical PM. Inc. Service Manger will veirfy if the consignee is capable of performing the rework.