RecallDepth

Elecsys Troponin T; catalog number 2017644

This recall has been terminated (originally issued May 29, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
April 25, 2003
Posted
May 29, 2003
Terminated
June 19, 2003
Recall Number
Z-0870-03
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.

Distribution

United States

Lot / Code Info

All lots.

Root Cause

Other

Action Taken

An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.

More recalls by Roche Diagnostics Corp.