RecallDepth

Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton.

This recall has been terminated (originally issued May 29, 2003).

Company
Dade Behring Inc. Rte 896, Glasgow Business Community
Recall Initiated
April 21, 2003
Posted
May 29, 2003
Terminated
July 31, 2003
Recall Number
Z-0873-03
Quantity
662 cartons
Firm Location
Newark, DE
Official Source
View on FDA website ↗

Reason for Recall

test results may be inaccurate by 0.2 to 0.5 ng/mL

Distribution

The product was shipped to hospitals and clinic laboratories nationwide. There are eight government/military accounts; one of which is foreign.

Lot / Code Info

Product code number RF451. Lot number CD4085 exp 3/25/04

Root Cause

Other

Action Taken

The recalling firm telephone each customer on 4/24-28/03 to determine if any customer used the lot (CD4085). If so, the customer was instructed to retest all samples with an alternate lot. All customers were instructed to discard any remaining inventory. As a follow up to the phone calls, the recalling firm issued a recall letter dated 4/25/03 informing the customer of the problem and to discard any remaining product.

More recalls by Dade Behring Inc. Rte 896, Glasgow Business Community

View all recalls by this company →