RecallDepth

Advanced D-Dimer. Fibrinogen and fibrin split products Catalog numbers OQWM11 and OQEM13.

This recall has been terminated (originally issued February 15, 2005).

Company
Dade Behring Inc. Rte 896, Glasgow Business Community
Recall Initiated
December 3, 2004
Posted
February 15, 2005
Terminated
October 18, 2005
Recall Number
Z-0538-05
Quantity
25381 units
Firm Location
Newark, DE
Official Source
View on FDA website ↗

Reason for Recall

decrease in product stability after reconstitution and reduced stability on board certain instrument systems

Distribution

The product was shipped nationwide to medical facilities.

Lot / Code Info

Lot numbers 544531 to 544548 and 544502 to 544504.

Root Cause

Other

Action Taken

The recalling firm issued a recall letter to their customers informing them of the problem and the need to use internal quality controls with the analyzers. The letter instructed customer to discard lot # 544545 and reduce the stability for the other lot numbers involved.