Boston Scientific LeVeen SuperSlim Needle Electrode, 2.0 cm/25 cm, sterile, Boston Scientific, Spencer, IN; REF 26-227, UPN M001262270. The product is used with radiofrequency generator for the thermal coagulation necrosis of soft tissues,including partial or complete ablation of nonresectable liver lesions.

This recall has been terminated (originally issued September 11, 2008).

Company
Boston Scientific Corp
Recall Initiated
April 30, 2008
Posted
September 11, 2008
Terminated
August 4, 2009
Recall Number
Z-1914-2008
Quantity
1,356
Firm Location
Spencer, IN
Official Source
View on FDA website ↗

Reason for Recall

The cannula may become detached from its correct orientation inside the handle and may prevent retraction of the tines.

Distribution

Nationwide, Argentina, Belgium, Canada, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Iraq, Ireland, Italy, Jordan, Korea, Lebanon, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, and United Arab Emirates.

Lot / Code Info

All lots (Lots 0007754222, 0007757352, 0007774377, 0007778866, 0007787319, 0007802363, 0007819373, 0008025066, 0008033462, 0008119925, 0008126720, 0008139108, 0008167709, 0008187490, 0008189883, 0008191619, 0008272777, 0008274731, 0008301843, 0008900347, 0009106193, 0009261704, 0009261705, 0009261706, 0009275205, 0009292249, 0009344591, 0009382115, 0009422549, 0009437649, 0009446512, 0009464930, 0009505685, 0009532051, 0009569176, 0009595712, 0009647788, 0009663657, 0009749621, 0009783669, 0009799393, 0009993406 0011311216 and 0011517938.)

Root Cause

Device Design

Action Taken

Consignees were notified by recall letter on 4/30/08, sent fexex priority mail, to cease use of the products and to return them to the firm. For assistance, contact Boston Scientific at 1-508-683-4186.