Animas IR 1250 Insulin Pump.

This recall has been terminated (originally issued June 25, 2005).

Company
Animas Corporation
Recall Initiated
April 11, 2005
Posted
June 25, 2005
Terminated
January 24, 2006
Recall Number
Z-0929-05
Quantity
2081 units
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

pump failed to respond to keypad button pressess after using the food database feature

Distribution

The products were shipped to distributors in FL, TX, UT, and VA. The products were also shipped to patients nationwide.

Lot / Code Info

Model IR1250

Root Cause

Other

Action Taken

The recalling firm issued a recall notice dated 4/11/05 to all patients and healthcare providers via first class mail. The letter instructs the patient to reset the pump, thus correcting any problems caused by the software bug, and to discontinue use of the food database. All distributors were contacted by phone on 4/12/05 and instructed to return all pumps in their inventory. On 4/25/05, the recalling began emailing and telephoning patients to arrange for replacement and return of the devices.