ProSpecT Campylobacter Microplate Assay. Catalog #2476096.
This recall has been terminated (originally issued August 17, 2004).
- Company
- Remel Inc
- Recall Initiated
- July 15, 2002
- Posted
- August 17, 2004
- Terminated
- December 28, 2006
- Recall Number
- Z-1372-04
- Quantity
- 382
- Firm Location
- Lenexa, KS
- Official Source
- View on FDA website ↗
Reason for Recall
A potential contamination was noted in the conjugate in ProSpecT Campylobacter Microplate Assay. The particulate in the conjugate may cause the dropper tip to become plugged.
Distribution
Entire lot was shipped to one customer in Russelsheim, Germany.
Lot / Code Info
Lot #910718
Root Cause
Other
Action Taken
Remel contacted the customer with a letter, requesting a count of the affected lot, and offered replacement of the kits.