ProSpecT Campylobacter Microplate Assay. Catalog #2476096.

This recall has been terminated (originally issued August 17, 2004).

Company
Remel Inc
Recall Initiated
July 15, 2002
Posted
August 17, 2004
Terminated
December 28, 2006
Recall Number
Z-1372-04
Quantity
382
Firm Location
Lenexa, KS
Official Source
View on FDA website ↗

Reason for Recall

A potential contamination was noted in the conjugate in ProSpecT Campylobacter Microplate Assay. The particulate in the conjugate may cause the dropper tip to become plugged.

Distribution

Entire lot was shipped to one customer in Russelsheim, Germany.

Lot / Code Info

Lot #910718

Root Cause

Other

Action Taken

Remel contacted the customer with a letter, requesting a count of the affected lot, and offered replacement of the kits.