1. CAMHB W/LHB, MANUAL, 11ML, 10/BOX YCP112-10 2. BRUCELLA BROTH, 11ML, 10/BOX YT3450
- Company
- Remel Inc
- Recall Initiated
- November 12, 2025
- Posted
- November 21, 2025
- Recall Number
- Z-0585-2026
- Quantity
- 1,000 US, 280 ROW
- Firm Location
- Lenexa, KS
Reason for Recall
Products may contain contamination, which may result in a darker or brown media color.
Distribution
Worldwide - US Nationwide distribution in the states of AK, CA, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, NE, NY, OH, OK, PA, RI, SD, TN, VA, WA, WI and the countries of Brazil, Canada, China, Cyprus, Finland, Great Britain, Ireland, Moldova, Norway, Singapore.
Lot / Code Info
1. UDI-DI 848838018831, lots 305529, 311228, 316019, 320993 2. UDI-DI 848838091353, lot 317287
Root Cause
Under Investigation by firm
Action Taken
On November 13, 2025, firm began notifying customers via Urgent Medical Device Recall letters. Customers were advised to review results and destroy any remaining inventory of affected lots. Affected product should not be used for any clinical laboratory testing. Customers should complete the acknowledgement form provided by the firm indicating number of units destroyed in order to receive replacement/credit.