Dinalink brand Communication Cable which is an accessory to a medical device Dinalike which is an isolated protocol converter used to concert the serial data from a DINAMAP Monitor to that of the APEXPRO telemetry transmitter.

This recall has been terminated (originally issued July 20, 2004).

Company
Ge Medical Systems Information Technologies
Recall Initiated
January 27, 2004
Posted
July 20, 2004
Terminated
October 5, 2009
Recall Number
Z-1015-04
Quantity
36
Firm Location
Tampa, FL
Official Source
View on FDA website ↗

Reason for Recall

The potential exists for external electrical noise to be processed as irregular pacing thereby masking potential asystole events.

Distribution

Product was distributed to 7 domestic hospitals in PA, SC, GA, FL, WI, SD AND WY.

Lot / Code Info

No serial numbers but the Dinalink protocol converter and cable has the product identifier of ''ILC1931''

Root Cause

Other

Action Taken

The firm mailed an 'Urgent Medical Device Correction' letter to consignees on January 27, 2004 advising them of the external electrical noise problem and informing them to not use the Dinalink ILC1931 with patients with pace makers until a fix or replacement can be provided. By 5/17/2004 corrective replacement cables had been provided to all consignees.