Power cord manufactured by ElectriCord Manufacturing Company , P/N 316579, used with DINAMAP Patient Monitors as Follows: DINAMAP PRO Series, DINAMAP PRO 1000 Monitor under both the Critikon and GE Medical Systems. The company name EliectriCord is on the power cord plug.

This recall has been terminated (originally issued October 15, 2003).

Company
Ge Medical Systems Information Technologies
Recall Initiated
April 30, 2003
Posted
October 15, 2003
Terminated
May 3, 2012
Recall Number
Z-0011-04
Quantity
16,339 Power Cords
Firm Location
Tampa, FL
Official Source
View on FDA website ↗

Reason for Recall

Reports of broken plug ground pins on AC power cords for patient monitors.

Distribution

The recalled AC power cords were distributed to approximately 2600 Domestic hospital accounts nationwide and 180 international accounts.

Lot / Code Info

Lot numbers 01-08 through 02-16 which are embossed into the power cord metal blades and also those with no lot numbers.

Root Cause

Component design/selection

Action Taken

The recalling firm mailed a Recall Notification Letter with response form and corretive power cords beginning on April 30, 2003 until August 14, 2003 as corrective cords became available. Consignees were instructed on how to determine if they have the recalled cords and to destroy any found by cutting the plug from the wire. Also corrective AC power cords were provided to each consignee.