propper brand cal-mark(R), Micro-Hematocrit Calibrated Capillary Tubes, CALIBRATED-HEPARINIZED, reorder no. 043021, 6 Vials, 100 Tubes per Vial, Length 75 mm., 0.55 mm. I.D., 1.40-1.75mm. O.D. Calibrated 60mm from end for use in direct reading centrifuges. Fill to the 60mm mark with blood and seal opposite end by flame or with sealing compound. If centrifuged on a direct reading centrifuge, read percent hematocrit directly from scale on the centrifuge. Otherwise, read in normal manner as a percent ratio of packed cells with 60mm of sample as 100%. Firm on label: propper manufacturing co., inc., long island city, n.y. 11101.

Company
Propper Mfg Co Inc
Recall Initiated
August 9, 2002
Posted
October 24, 2003
Terminated
October 23, 2003
Recall Number
Z-0051-04
Quantity
178 boxes
Firm Location
Long Island City, NY

Reason for Recall

Deviation in capillary tubes could cause a 10% increase in read out values which could make a slight anemic patient appear to have a normal reading.

Distribution

Distributors located in TX, UT, ID, IL, IA, IN, MN, WI, OH, CT, MO, PA, VA.

Lot / Code Info

Reorder No. 043021, Lot # C0402

Root Cause

Other

Action Taken

On 8/9/02, Propper faxed a 'Request for Return' of the product to the consignees advising them of the problem and that a replacement will be sent for all returned product. On 11/27/02, Propper faxed a second 'Request for Return of Product' to its consignees. The firm destroyed the recovered and held stock.