RecallDepth

Propper short gas-chex EO sterilization indicators, 250 strips per box, Reorder No. 269001, Propper Manufacturing Co., Inc., Long Island City, N.Y. 11101

Company
Propper Mfg Co Inc
Recall Initiated
February 7, 2008
Posted
September 20, 2008
Terminated
August 3, 2012
Recall Number
Z-2330-2008
Quantity
1757 cases (7030 boxes)
Firm Location
Long Island City, NY

Reason for Recall

The firm discovered a production specification discrepancy that may cause the indicators to show an inaccurate result.

Distribution

Nationwide and Puerto Rico

Lot / Code Info

Reorder No. 269001; Lots 6081, 6043, 6021, 7042, and 7013

Root Cause

Equipment maintenance

Action Taken

On 2/07/08, Propper Mfg. Co. sent URGENT: DEVICE RECALL letters (dated 2/07/08) to the direct consignees, informing them of the recall of the affected lots of Gas-Chex EO Indicator Strips. The letters also instructed them to: (1) check their inventory to determine if they have the recalled lots in stock; (2) discontinue use of the affected lots; (3) reprocess any unused packs that have been processed with the affected lots of Gas-Chex EO Indicator Strips with non-recalled lots or an alternative product from another supplier; (4) return the affected lots to Propper Manufacturing; (5) complete the enclosed form and return via fax to Propper Mfg. Propper Mfg. will send a postage-paid label and instruction for return of the product, and will ship replacement product as soon as possible. Propper also recommended that consignees review the risk associated with items that have been monitored with the subject chemical indicators although a sterilization cycle failure is unlikely due to equipment controls and other monitoring methods. For questions, please contact Propper at 1-800-832-4300 during regular business hours (9 am to 4:30 pm).

More recalls by Propper Mfg Co Inc