RecallDepth

Leica M530 OHX Microscope System Supplied with 100 - 120 Vac.

This recall has been terminated (originally issued June 25, 2018).

Company
Leica Microsystems, Inc.
Recall Initiated
June 25, 2018
Terminated
February 26, 2019
Recall Number
Z-2659-2018
Quantity
123 devices total
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Recall is due to a design weakness of the power insert module.

Distribution

United States distribution: AR, AZ, CA, CO, CT, DE, FL, IL, IN, KS, KY, LA, MO, NJ, NY, NV, OH, OR, PA, TX, UT, PR. International distribution to Japan, Brazil, Columbia, Ecuador, Mexico, Taiwan

Lot / Code Info

US: 90116001, 30118001, 51215001, 201117001, 30916001, 270216001, 100917001, 31117001, 00817001, 0316001, 31217001, 180118001, 200218001, 151217001, 21217001, 21214001, 80218001, 170118001, 01017002, 101017001, 170915001, 150117001, 150217001, 291117001, 260517001, 300915001, 0318001, 30318001, 100915001, 310717001, 140716001, 50516001, 281116001, 300517001, 0316001, 30115001, 200617001, 190117001, 170117001, 241217001, 261217001, 271217001, 90118001, 0617001, 250118001, 280115001, 191015001, 81117001, 150617001, 120417001, 140917001, 10318001, 140417001, 250717001, 80518001, 160616001, 170417001, 150417001, 11116001, 91017001, 81117002, 281117001, 20218001, 300118001, 131015001, 140318001, 110516001, 00318001, 160118001, 30518001, 200316001, 120917001, 210118001, 310717001, 30218001, 1217001, 190617001, 130617001, 280817001, 80317001, 310817001, 10917001, 60317001, 60218001, 80917001 230318001, 170318001;   Puerto Rico: 241216001, 261216001;   JAPAN: 140717001,171017001, 211117001, 291217001, 290118001, 210218001, 280218001, 020318001, 120318001, 160318001;   BRAZIL: 120416001, 220917001, 311017001, 70118001, 70318001;   COLOMBIA: 60517001;   ECUADOR: 170316001, 231216001, 190917001;   MEXICO: 31215001, 130117001, 91217001;   TAIWAN: 190815001, 220317001, 190317001, 40517001, 210717001, 210717002, 211217001

Root Cause

Device Design

Action Taken

Leica Microsystems notified customers on about 06/25/2018 via URGENT FIELD SAFETY NOTICE letter. Instructions included advising customers of the issue, informing them that Leica Microsystems representative will contact them to arrange the fiel update to the device, distribute the recall notification to all individuals that should be aware of the recall or if the device was further distributed and a response form was provided to complete and return. The notification letter also instructed customers to report any anomalies with the system to Leica Microsystems if they occur prior to the correction being made and to ensure a functional check is performed prior to each surgery as described in the user manual.

More recalls by Leica Microsystems, Inc.

View all recalls by this company →