Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.

This recall has been terminated (originally issued October 15, 2003).

Company
Roche Diagnostics Corp.
Recall Initiated
May 8, 2003
Posted
October 15, 2003
Terminated
December 9, 2003
Recall Number
Z-0012-04
Quantity
478
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.

Distribution

United States

Lot / Code Info

All units with current software versions (v 06-06 and 07-01).

Root Cause

Other

Action Taken

Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.