RecallDepth

FilmArray NGDS Warrior Panel NGDS-ASY-0007 is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system.

This recall is currently active, issued September 19, 2025. It was issued by Biofire Defense.

Company
Biofire Defense
Recall Initiated
July 10, 2025
Posted
September 19, 2025
Recall Number
Z-2609-2025
Quantity
130 Kits
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

FilmArray NGDS Warrior Panel (NGDS-ASY-0007) has an increased risk of internal control failures when testing positive blood culture, which may lead to delayed results and additional diagnostic workup.

Distribution

US Nationwide distribution in the states of AL, AZ, UT, CA, TX, DC , VA, FL, MD, OH, HI, NY, KY, LA, MA, MI, MN, MO, NC, NH, RI, NV, TN, IA, WA.

Lot / Code Info

UDI-DI: 00851458005136. Lot(Expiration): 324324D(2026-02-19), D241022(2026-01-22), 319524D(2025-11-20), 315424D(2025-10-09), 315224D(2025-10-03), 315324D(2025-10-08)

Root Cause

Component design/selection

Action Taken

On 7/10/2025, correction notices were mailed and emailed to customers who were asked to do the following: - If testing with only whole blood and/or sputum, there is no impact to the product s performance. Continue using panel kits as planned. - If testing with positive blood culture may occur, discontinue use, dispose of impacted kit(s) according to your lab s procedures, and only use replacement kits provided, or use impacted lots only with whole blood and/or sputum sample types. - Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties who may use this product, including others to whom you may have transferred our product. - Complete and return the acknowledgment of receipt form via email to support@biofiredefense.com

More recalls by Biofire Defense

View all recalls by this company →