JBAIDS Instrument System Software version 3.5.0.72. The JBAIDS instrument is a portable thermocycler and real-time fluorimeter capable of simultaneous and rapid identification of multiple biological warfare and clinical pathogens from a variety of biological specimens. The software and instrument provides an integrated environment for performing the required functions to control JBAIDS instruments and to perform PCR tests and analysis.

This recall has been terminated (originally issued September 15, 2014).

Company
Biofire Defense
Recall Initiated
August 13, 2014
Posted
September 15, 2014
Terminated
October 24, 2016
Recall Number
Z-2644-2014
Quantity
43 Units w/ laptops
Firm Location
Salt Lake City, UT
Official Source
View on FDA website ↗

Reason for Recall

BioFire Defense, LLC is initiating a field action for JBAIDS Software version 3.5.0.72 because the wrong version software may have been installed.

Distribution

Distributed USA (nationwide) and Guam.

Lot / Code Info

Model Number(s): JRPD-SFW-0001

Root Cause

Software Manufacturing/Software Deployment

Action Taken

Consignees were notified on 8/13/14 via letter and telephone and were provider with instructions on how to install the correct software.