2.5 V replacement rechargeable batteries for Hillrom WELCH ALLYN PocketScope handles, Product Code 72600

Company
Baxter Healthcare Corporation
Recall Initiated
May 9, 2025
Posted
June 11, 2025
Recall Number
Z-1962-2025
Quantity
277 units
Firm Location
Deerfield, IL

Reason for Recall

The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI/DI 00732094139488, Lot Number 25-043, Exp. August 2027

Root Cause

Under Investigation by firm

Action Taken

Baxter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 05/09/2025 via USPS First Class mail. The notice explained the issue, hazard involved, and requested to cease all use of the device and return it to Baxter. Please contact Baxter Technical Support to arrange for return and replacement of your affected battery or Welch Allyn PocketScope handle at 800-535-6663 between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday. Press option 2, then select option 1 or 2 for language preference, then press option 1 for physical exam, and then press 3 for physical exam products. Forward a copy of the notice to any other departments with the institution that may use the affected product. Distributors were directed to notify their consignees.