Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system

This recall is currently active, issued June 20, 2025. It was issued by Barco N.V. President Kennedypark 35 Kortrijk Belgium.

Company
Barco N.V. President Kennedypark 35 Kortrijk Belgium
Recall Initiated
May 9, 2025
Posted
June 20, 2025
Recall Number
Z-1994-2025
Quantity
221
Official Source
View on FDA website ↗

Reason for Recall

Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures

Distribution

US Nationwide distribution in the states of TX, OH.

Lot / Code Info

REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726; K9303300/05415334002180/2531537815, 2531557845, 2531557866, 2531558588, 2531558727, 2531558738, 2531558753, 2531558754, 2531558759, 2531561761, 2531561785, 2531572541, 2531572557, 2531572568, 2531572576, 2531572847, 2531572862, 2531572872, 2531633606, 2531633662; K9303311/05415334026810/2531537674, 2531572141, 2531580810, 2531580935, 2531582319, 2531582331, 2531582341, 2531588386, 2531588401, 2531588404, 2531588456, 2531591560, 2531592036, 2531592037, 2531592046, 2531593872, 2531594457, 2531594475, 2531594481, 2531594503, 2531594515, 2531597664, 2531597713, 2531597717, 2531597726, 2531597795, 2531597806, 2531597809, 2531597810, 2531597820, 2531599836, 2531605356, 2531605390; K9306210/2531542260, 2531542263, 2531542270, 2531542292, 2531543553, 2531543568, 2531543588, 2531582676, 2531582691, 2531588558, 2531588562, 2531593741, 2531593749, 2531593752, 2531593763, 2531593764, 2531593768, 2531593777, 2531593784, 2531595829, 2531595831, 2531601786; K9306220/ 2531540107, 2531542441, 2531542470, 2531543538, 2531543549, 2531543611, 2531543613, 2531543640, 2531563132, 2531572740, 2531572759, 2531572770, 2531572814, 2531572827, 2531575652, 2531594331, 2531594332, 2531594336, 2531594348, 2531594350; K9306410/2531572228, 2531582713, 2531582724, 2531582725, 2531584168, 2531584176, 2531584191, 2531588591, 2531588593, 2531588598, 2531588616, 2531588626, 2531604519, 2531604535, 2531604544, 2531604545, 2531604549, 2531604550, 2531604551, 2531604593, 2531604637, 2531604649, 2531604658; K9306420/ 2531537844, 2531537870, 2531537889, 2531537901, 2531537902, 2531556037, 2531556043, 2531572419, 2531572433, 2531572440, 2531595759

Root Cause

Nonconforming Material/Component

Action Taken

On 5/9/2025, recall notices started being distributed to customers informing them of the following: 1) Assess & locate the non-installed affected units and contain them from further distribution. B) Assess & locate affected units that are installed at your customer side. Please make your customers aware of this notification. C) The impacted units will need to be sent back. Further information on the swaps will be provided in the coming days. D) Fill in and return the Acknowledgement Letter. If you don t have the Key Account Manager s contact details, or need other support, you can contact the firm 24/7 via their Support web page: https://www.barco.com/en/support