Barco MDNC-3421 DISPLAY, DIAGNOSTIC RADIOLOGY
This recall has been terminated (originally issued January 22, 2021).
- Recall Initiated
- January 22, 2021
- Terminated
- November 18, 2022
- Recall Number
- Z-1172-2021
- Quantity
- 1 unit
- Official Source
- View on FDA website ↗
Reason for Recall
A number of MDNC-3421 displays were shipped while not being DICOM-compliant. These displays will not pass DICOM acceptance tests.
Distribution
WI.
Lot / Code Info
Model Number: MDNC-3421, Serial number 2590263533, UDI/DI 05415334028388.
Root Cause
Process change control
Action Taken
On January 22, 2021 the firm notified the importer in the USA, Barco Inc.