a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
This recall is currently active, issued October 25, 2024. It was issued by Baxter Healthcare Corporation.
- Company
- Baxter Healthcare Corporation
- Recall Initiated
- September 18, 2024
- Posted
- October 25, 2024
- Recall Number
- Z-0228-2025
- Quantity
- 2,967 beds
- Firm Location
- Deerfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Distribution
Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.
Lot / Code Info
All serial numbers with software version 1.27.000. UDI-DI 00887761985162: Product codes P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. UDI-DI 00887761999961: Product codes P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.
Root Cause
Software design
Action Taken
The recalling firm issued letters on and dated 9/18/2024 via first class mail. The letter informed the consignee of the issue, the beds involved, and that Baxter will be upgrading the software on all Centrella Smart+ beds with the affected software version to resolve the issue. An Attachment was enclosed listing the affected product codes for the beds. The hazard involved and the actions to be taken by customers were provided, which included: 1. To mitigate the patient risk associated with this software issue, customers using the Rauland Responder 5 Nurse Call system should ensure the bed exit alert is silenced at the bedside first before silencing the alert at the in-room nurse call wall unit. 2. Please acknowledge receipt of this notification by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com. Log in to the portal using the account number listed in the enclosed reply form instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 3. If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them.