RecallDepth

MEDLINE STERILE IRIS SCISSORS CVD/STD, REF DYNJ04049

This recall is currently active, issued September 18, 2023. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
August 3, 2023
Posted
September 18, 2023
Recall Number
Z-2595-2023
Quantity
324,451 devices
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

4.5" Iris Scissors, Sterile, Curved (DYNJ04049) are being recalled due to the potential of the tip protector to fall off. The sharp tip of the scissors could compromise the sterile barrier system by puncturing large and/or small holes that may not be visible to the user.

Distribution

Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Canada, Hong Kong, Taiwan, Singapore

Lot / Code Info

UDI/DI 10080196847273, all lots

Root Cause

Under Investigation by firm

Action Taken

Medline Industries issued a recall notice to its consignees by email on 08/03/2023. The notice explained the problem with the device, the potential hazard, and requested the following: Immediately check your stock for the affected item number and the affected lot numbers listed on the recall portal. Destroy affected product. Distributors or those who resold or transferred the product were directed to notify their customers and have them destroy any affected product. If you have any questions, please contact the Recall Department at 866-359-1704.

More recalls by Medline Industries, Lp - Northfield

View all recalls by this company →