RecallDepth

JOURNEY II BCS ARTICULAR INSERT, LEFT 10 MM SIZE 5-6, REF 74027262; knee prosthesis

This recall is currently active, issued October 24, 2023. It was issued by Smith & Nephew Inc.

Company
Smith & Nephew Inc
Recall Initiated
September 27, 2023
Posted
October 24, 2023
Recall Number
Z-0157-2024
Quantity
28 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

The JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM. And also, a JRNY II BCS CNSTRD ART ISRT 5-6 LT 10MM was laser etched, labeled and packaged as a JRNY II BCS XLPE ART ISRT SZ 5-6 LT 10MM.

Distribution

Worldwide distribution - US Nationwide and the country of Japan.

Lot / Code Info

UDI/DI 00885556178010, Batch Number 22BM18348

Root Cause

Component change control

Action Taken

Smith & Nephew, Inc. issued an Urgent Medical Device Recall Notice to its consignees on 09/27/2023 via email and overnight mail. The notice explain the issue, potential risk, and requested the following actions: Inspect inventory, locate, and quarantine the affected devices. Sales representatives, district offices, or distributors were directed to notify their customers The firm is seeking return of unused products.

More recalls by Smith & Nephew Inc

View all recalls by this company →