Prismaflex M Set: M60 (product code 106696), M100 (product code 106697). For use in providing continuous fluid management and renal replacement therapies.

Company
Baxter Healthcare Corporation
Recall Initiated
September 14, 2022
Posted
October 12, 2022
Recall Number
Z-0081-2023
Quantity
524,376 devices
Firm Location
Deerfield, IL

Reason for Recall

Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.

Distribution

US Nationwide

Lot / Code Info

UDI: 07332414064549 (product code 106696), 07332414064556 (product code 106697); All lot codes

Root Cause

Error in labeling

Action Taken

Baxter sent an Urgent Medical Device Correction notice to its direct consignees by USPS on 09/14/2022. The notice explained the Estonian translation issue and requested that the following actions be taken by consumers: 1. Use of the Prismaflex HF20 set should be restricted to patients with a body weight greater than 8kg (18lb). Use of the Prismaflex M60 and ST60 sets should be restricted to patients with a body weight greater than 11kg (24lb). Use of the Prismaflex M100, ST100, M150, ST150, HF1000 and HF1400 sets should be restricted to patients with a body weight greater than 30kg (66lb). 2. Customers who are not referring to the Estonian IFU should continue to follow the IFU instructions in their language. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com/. 4. If you purchased this product from a distributor, please note that responding via the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal.

More recalls by Baxter Healthcare Corporation

View all recalls by this company →