RecallDepth

smith&nephew EVOS MEDIAL DISTAL FEMUR PLATE, R, 5H, 90MM, REF 72573201

Company
Smith & Nephew Inc
Recall Initiated
November 1, 2021
Recall Number
Z-0364-2022
Quantity
35 units
Firm Location
Memphis, TN

Reason for Recall

Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.

Distribution

US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.

Lot / Code Info

Batch Number: 20HM10255, 20HM10256, 20HM10257, UDI

Root Cause

Under Investigation by firm

Action Taken

The firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.

More recalls by Smith & Nephew Inc