RecallDepth

epoc BGEM Test Card 25 Test Cards/box. In vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood Product Siemens SMN#/Model #: 10736515

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
June 1, 2022
Recall Number
Z-1372-2022
Quantity
263 Boxes
Firm Location
Norwood, MA

Reason for Recall

Discrepant (low bias) pO2 results, could be an incorrect diagnosis of hypoxemia and initiation of unnecessary or inappropriate respiratory support.

Distribution

AL CA CO FL IA ID KS NM OK TX

Lot / Code Info

UDI-DI: 00809708121860 Lot Number: 11-22005-50  Exp. Date: 22-June-2022

Root Cause

Under Investigation by firm

Action Taken

Siemens has issued an Urgent Medical Device Correction letter POC 22-010.A.US via FedEx beginning on June 1, 2022. An Urgent Field Safety Notice (UFSN) was issued to OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities. Letter states reason for recall, health risk and action to take: Please dispose any inventory of epoc BGEM test card lot 11-22005-50 currently in your possession in accordance with local and state disposal requirements. You may request free of charge replacement product from your local Siemens or distributor office. Please review your inventory of this product and assess your laboratory s replacement needs. Please use an epoc BGEM test card from another unaffected lot. Review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check and Product Replacement Form attached to this letter. Please retain this letter with your laboratory records and forward it to those who may have received this product. Questions contact your Siemens Customer Care Center or your local Siemens technical support representative.

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →