Olympus Uretero-reno videoscope, Model No. URF-V2

Company
Olympus Corporation Of The Americas
Recall Initiated
March 25, 2022
Recall Number
Z-1116-2022
Quantity
1
Firm Location
Center Valley, PA

Reason for Recall

The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.

Distribution

US Nationwide distribution in the state of AZ.

Lot / Code Info

1 unit, serial number 2825505F UDI 04953170343582

Root Cause

Process change control

Action Taken

On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.