Olympus Uretero-reno videoscope, Model No. URF-V2

This recall is currently active, issued March 25, 2022. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
March 25, 2022
Recall Number
Z-1116-2022
Quantity
1
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.

Distribution

US Nationwide distribution in the state of AZ.

Lot / Code Info

1 unit, serial number 2825505F UDI 04953170343582

Root Cause

Process change control

Action Taken

On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.