Olympus Uretero-reno videoscope, Model No. URF-V2
- Recall Initiated
- March 25, 2022
- Recall Number
- Z-1116-2022
- Quantity
- 1
- Firm Location
- Center Valley, PA
Reason for Recall
The bending section of the URF-V2 may crack, resulting in an abnormal shape or abnormal angulation of the bending section.
Distribution
US Nationwide distribution in the state of AZ.
Lot / Code Info
1 unit, serial number 2825505F UDI 04953170343582
Root Cause
Process change control
Action Taken
On March 25, 2022, the firm contacted the affected consignee to inform them that certain manufacturing changes by a supplier led to the possibility that a part within the bending section may crack. Olympus repaired the affected URF-V2 unit free of charge.