RecallDepth

Medline procedural kits labeled as: ARTHROGRAPH TRAY, Pack Number MNS12670A

This recall is currently active, issued August 28, 2024. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
April 8, 2024
Posted
August 28, 2024
Recall Number
Z-2754-2024
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Distribution

Worldwide

Lot / Code Info

MNS12670A, Lot Number 2022112990

Root Cause

Under Investigation by firm

Action Taken

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

More recalls by Medline Industries, Lp - Northfield

View all recalls by this company →