RecallDepth

Stratus CS Acute Care cTNI TestPak, SMN 10445071 Product Usage: In vitro diagnostic test for the measurement of cardiac troponin I in heparinized plasma. Cardiac troponin I measurements can be used as an aid in the diagnosis of acute myocardial infarction (AMI).

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
May 29, 2019
Terminated
May 12, 2022
Recall Number
Z-2135-2019
Quantity
2312
Firm Location
Norwood, MA

Reason for Recall

Potential for increased rate of false positive results and/or increased rate of random non-repeatable false positive results.

Distribution

US Nationwide Distribution

Lot / Code Info

Lot # 238330002 forward Expiration Dates - May 25, 2019 - November 2, 2019

Root Cause

Under Investigation by firm

Action Taken

On May 29, 2019, Siemens issued Urgent Medical Device Correction notices to U.S.A. customers via FedEx courier service and Urgent Field Safety Notices to international customers via email. Actions to be Taken by the Customer: 1. Do not use TestPaks within 2 mos of expiration. 2. Repeat samples with cTNL results above 0.07 ng/mL or your institution's established 99th percentile value. 3. Complete and return the Field Correction Effectiveness Check form 4. Retain a copy of the letter with your laboratory records and forward a copy of the notice to anyone to whom you may have distributed the product. Siemens is working to resolve this issue and will contact customers when additional information is available. Siemens will reimburse users for repeat tests and discarded cTNL TestPaks within 2 months of the printed expiration date associated with this field corrective action.

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →