Roche / Hitachi MODULAR Analyzer Systems-Immunoassay Analyzer-Model/Catalog/Part Number: 11568248001/11568248692 - Elecsys 2010 analyzer (disk system) and 11804014001/11804014692 - Elecsys 2010 analyzer (rack system) Medical Device Listing number: D053244 - Elecsys 2010 Elecsys analyzer is a fully automated, random access, computer controlled analytical systems for quantitative and qualitative determinations of analytes in body fluids.

This recall has been terminated (originally issued July 5, 2017).

Company
Roche Diagnostics Corporation
Recall Initiated
July 5, 2017
Terminated
April 30, 2019
Recall Number
Z-0499-2018
Quantity
255 instruments
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A software malfunction can occur on the cobas e 411 and Elecsys 2010 analyzers in the Sample& Control data file which may lead to a potential data mismatch.

Distribution

Nationwide Distribution

Lot / Code Info

none

Root Cause

Software design

Action Taken

All consignees were notified by Urgent Medical Device Correction sent via UPS Ground (receipt required) on July 5, 2017. The customer is advised to perform the following actions: cobas e 411: Until the software update is available, use the workaround outlined in the Urgent Medical Device Correction (UMDC). Elecsys 2010: Use the workaround outlined in this Urgent Medical Device Correction (UMDC). Non-responding accounts will be defined as users for which confirmation of receipt of the Urgent Medical Device Correction (UMDC) cannot be obtained. These accounts will be monitored on an ongoing basis and follow-up attempts will be made accordingly. Workaround for customers: he customer is advised to perform the following actions: cobas e 411: Until the software update is available, use the workaround outlined in the Urgent Medical Device Correction (UMDC). Elecsys 2010: Use the workaround outlined in this Urgent Medical Device Correction (UMDC). For further questions, please call (317) 521-7259

More recalls by Roche Diagnostics Corporation

View all recalls by this company →