Covidien Nellcor SpO2 Forehead Sensor Item Code: RS10; The Covidien Nellcor Forehead SpO2Sensor, model RS10, is indicated for use when continuous non-invasive arterial oxygen saturation and pulse rate monitoring are required for patients weighing more than 40 kg.

This recall has been terminated (originally issued December 16, 2016).

Company
Covidien Llc
Recall Initiated
December 16, 2016
Terminated
December 4, 2018
Recall Number
Z-1125-2017
Quantity
7453 Units
Firm Location
North Haven, CT
Official Source
View on FDA website ↗

Reason for Recall

Labeling error: The label indicates the product is latex free, but the headband component included with the forehead sensor does contain latex.

Distribution

Nationwide. Foreign: Belgium Germany France Netherlands

Lot / Code Info

Lot Numbers: 161510095H, 161620089H, 161690203H, 161760132H, 161900091H,161970090H, 162040107H, 162110103H, 162320191H, 162390212H

Root Cause

Process control

Action Taken

Medtronic initiated on December 16, 2016. Customers were notified via Federal Express and the letter instructed customers to discontinue use of the product and return all inventory to Medtronic (Covidien) 195 McDermott Road, North Haven, CT 06473 Attn: Field Returns Department. Complete a Recalled Product Return Form (Attachment A, included with letter) and submit it to customer service so a Returned Goods Authorization number (RGA#) can be issued. Questions contact Medtronic representative or customer service at 800-882-5878.