Infinite 200 PRO The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conforming to the ANSI/SBS standards.

This recall has been terminated (originally issued August 25, 2016).

Company
Tecan Us, Inc.
Recall Initiated
July 25, 2016
Posted
August 25, 2016
Terminated
January 23, 2017
Recall Number
Z-2663-2016
Quantity
27 units
Firm Location
Morrisville, NC
Official Source
View on FDA website ↗

Reason for Recall

The batteries have the potential to overheating, melt or char.

Distribution

Worldwide Distribution - US including AK, AL, AR, CA, FL, GA, HI, KY, MA, MD, NC, NY, PA, PR, SC, TN, TX, UT, VA. International account: Canada.

Lot / Code Info

Instrument Material #30050303, PC Serial numbers: CND5219H0B, Instrument Serial #1208002673; PC Serial numbers: CND44273RW, Instrument Serial #1211001074; PC Serial numbers: CND44273SC, Instrument Serial #1309006773; PC Serial numbers: 2CE3461RNR, Instrument Serial #1311008588; PC Serial numbers: 2CE33625KT, Instrument Serial #1401006356; PC Serial numbers: 2CE33626Q9, Instrument Serial #1403000833; PC Serial numbers: 2CE3461RR3, Instrument Serial #1403003928; PC Serial numbers: 2CE33625M4, Instrument Serial #1403005482; PC Serial numbers: 2CE41503JF, Instrument Serial #1405000420; PC Serial numbers: CND5219H4R, Instrument Serial #1405002510; PC Serial numbers: 2CE42406YJ, Instrument Serial #1406009472; PC Serial numbers: 2CE4220LGP, Instrument Serial #1410002321; PC Serial numbers: 2CE424072G, Instrument Serial #1410004940: PC Serial numbers: 2CE4240DJ7, Instrument Serial #1411008631; PC Serial numbers: 2CE424071M, Instrument Serial #1412005274; PC Serial numbers: CND44273XR, Instrument Serial #1503001536; PC Serial numbers: CND5219H1S, Instrument Serial #1503002986; PC Serial numbers: CND44262KJ, Instrument Serial #1503004890; PC Serial numbers: CND442625D, Instrument Serial #1501008506; PC Serial numbers: CND442627J, Instrument Serial #1501008513; PC Serial numbers: CND44273SB, Instrument Serial #1505009092; PC Serial numbers: CND442626B, Instrument Serial #1509001022; PC Serial numbers: CND5219H11, Instrument Serial #1509008146; PC Serial numbers: CND5219H6D, Instrument Serial #1511009716; PC Serial numbers: 2CE3461RSX, Instrument Serial #1512005211; PC Serial numbers: 2CE3381HT1, Instrument Serial #1601005141 and PC Serial numbers: CND5219H2K, Instrument Serial #1510004918.

Root Cause

Nonconforming Material/Component

Action Taken

Tecan contacted all customers by Safety Announcement, via certified mail starting July 25, 2016. The letter identified the product, the problem, and the action to be taken by the customer. Because affected batteries pose a fire and burn hazard, it is extremely important that customers check whether their battery is affected. Customers were advised to go immediately to the HP Battery Recall website at: http://www.HP.com/go/batteryprogram2016 to check if their battery is affected by this recall. Customers were also advised to complete and return the Customer Response Form at QA@tecan.com or fax it to 919-287-2961. Customers with questions should contact HP at 1-888-202-4320 or Tecan Technical Service's Helpdesk at 800-832-2687, option 1.