IntelliVue MX800 patient monitor, model no. 865240

This recall has been terminated (originally issued June 30, 2020).

Company
Philips North America, Llc
Recall Initiated
June 30, 2020
Terminated
January 25, 2024
Recall Number
Z-2696-2020
Quantity
28
Firm Location
Andover, MA
Official Source
View on FDA website ↗

Reason for Recall

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.

Distribution

Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.

Lot / Code Info

Serial numbers  DE50887321 DE50887434 DE50887128 DE50887025 DE50887196 DE50887429 DE50887125 DE50887043 DE50887303 DE50887420 DE50887106 DE50887164 DE50887322 DE50887607 DE50887061 DE50887179 DE50887425 DE50887427 DE50887142  DE50887446 DE50887152 DE50887064  DE50887445 DE50887137 DE50887063

Root Cause

Process control

Action Taken

On June 30, 2020, the firm sent Urgent Medical Device Correction letters to affected customers. Customers were instructed to always ensure that the monitor is grounded during operation with mains power. Affected products should be removed from use until the power supply of the monitor has been replaced. Philips Customer Service will contact customers when a replacement power supply is ready for installation. Philips expects all replacements to be complete within six months. If you need any further information or support concerning this issue, please contact your local Philips representative or Philips Customer Support at 1-800- 722-9377.