Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body.

Company
Philips North America, Llc
Recall Initiated
November 19, 2020
Terminated
April 19, 2023
Recall Number
Z-0852-2021
Quantity
2714
Firm Location
Andover, MA

Reason for Recall

A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.

Distribution

Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, Algeria, Argentina, Australia, Bahrain, Austria, Belgium, Brazil, China, Columbia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Guatemala, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Kuwait, Latvia, Libya, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, United Kingdom, Venezuela.

Lot / Code Info

Model# 882310,882311,882313 All Lots

Root Cause

Device Design

Action Taken

On 11/23/2020, Philips issued Urgent Device Correction notices to customers via press release, letter, telefax, telephone, e-mail, visit, etc. Due to a software defect Philips is notifying all affected users of the JETStream Workspace Thyroid Analysis application recommending that customers calculate thyroid uptake utilizing the method provided in the notices and not using the Transfer Coefficient parameter.