MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.

Company
Roche Molecular Systems, Inc.
Recall Initiated
June 3, 2020
Terminated
October 4, 2023
Recall Number
Z-2551-2020
Quantity
917
Firm Location
Branchburg, NJ

Reason for Recall

When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.

Distribution

Worldwide distribution - US Nationwide distribution.

Lot / Code Info

All units when used with Sample Transfer protocol version 3.0

Root Cause

Software change control

Action Taken

On June 5, 2020, the firm distributed Urgent Medical Device Correction letters to consignee accounts. The UMDC instructs the customers to determine whether or not their laboratory is using the affected Sample Transfer protocol version 3.0 with their MagNA Pure 96 instrument. If the affected protocol is found on the instruments Control Unit, it needs to be immediately deleted by following the instructions provided in the UMDC. A newly updated Sample Transfer protocol version 4.0 is available. Roche Affiliate organizations will provide instructions for obtaining the new version of the protocol. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about this recall.