cobas 5800 instrument, Material No. 08707464001. Used for automated Polymerase Chain Reaction (PCR) based Nucleic Acid Testing (NAT) in laboratory settings.

Company
Roche Molecular Systems, Inc.
Recall Initiated
September 8, 2022
Posted
October 17, 2022
Terminated
November 7, 2024
Recall Number
Z-0097-2023
Quantity
179 OUS
Firm Location
Branchburg, NJ

Reason for Recall

Reported false positive and invalid results on the affected devices due to anomalous baselines. This may lead to erroneous or delayed diagnoses.

Distribution

International distribution in the countries of Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Italy, Lithuania, Poland, Portugal, Spain, Sweden, Switzerland, UK, Chile, Colombia, Hong Kong, Jamaica, Japan, Nicaragua, Oman, Pakistan, UAE.

Lot / Code Info

Device Identifier: 07613336170076 All serial IDs are affected.

Root Cause

Under Investigation by firm

Action Taken

On September 8, 2022, the firm notified all affected customers via an Urgent Field Safety Notice letter. Customers were informed that the issue was attributed to WAKO lenses with a deposit on their surfaces, which can affect the signal in the detection unit. Roche will retrieve and analyze customer's cobas 5800 run data. If the evaluation determines that a lens cleaning is warranted, a service visit will be scheduled. If a lens cleaning is not warranted, no further actions are required. Customers should follow laboratory standard operating procedures to investigate the potential for false positive results for assays where a change in result reporting could impact patient management. In the case of testing for respiratory viruses or blood screening, uncovering false positive results more than one week old would be unlikely to change patient management. In the case of assays used to manage chronic diseases (e.g. hepatitis C) or those used in serial monitoring (e.g. HIV), only the most recent result for a patient would have the potential to affect management. Contact your local affiliate organization if there is any allegation of invalid or false positive results with the cobas 5800 system assays. This device is not currently available in the US.