RecallDepth

smith&nephew R3 MULTI HOLE HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Sizes REF/product number 48MM 71338663 50MM 71338664 52MM 71338665 54MM 71338666 56MM 71338667 58MM 71338668 Orthopedic implant component.

This recall is currently active, issued April 22, 2020. It was issued by Smith & Nephew Inc.

Company
Smith & Nephew Inc
Recall Initiated
April 22, 2020
Recall Number
Z-2335-2020
Quantity
1766 devices
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

A manufacturing error resulted in out of specification R3 Acetabular Shells

Distribution

Worldwide

Lot / Code Info

71338663, Lot Numbers: 18EM10564, 18EM23226, 18EM23226A, 18HM14882, 18JM12818, 18JM12819, 18JM12819A, 18KM09644, 18LM19836, 18MM07583, 19BM10203, 19CM19574, 19CM19582, 19DM28076, 18KM25982, 18MM07580, 19DM15683, 19EM19652, 19EM19655, 19EM19656, 19EM19657, 18KM25982A, 18GM18693, 18GM18700, 18HM04587, 18HM04588, 18HM04593, 18HM14887, 18JM02705, 18GM18693A, 18GM18700A, 18HM04588A, 18HM14887A  71338664, Lot Numbers: 18JM00846, 18JM00851, 18JM03933, 18JM03964, 18KM11934, 18KM14029, 18LM19837, 18LM21024, 19AM23671, 19AM23674, 19BM12881, 19DM15685, 19DM15687, 19EM19658, 19EM19659, 19EM19660, 18GM04676, 18GM04678, 18GM04680, 18HM14311, 18EM01929, 18EM14265, 18HM05120, 18JM03940, 19DM15688, 19DM28083, 19DM28084, 19DM28085, 19DM28086, 19DM28078, 19DM28079, 19EM19661, 19EM19662, 18JM03940A, 19DM28083A, 18KM25957, 18KM25958, 19AM05159, 19AM05159A  71338665, Lot Numbers: 18GM06572, 18GM07678, 18HM16467, 18JM01340, 18JM01342, 18JM08808, 18JM13325, 18JM17690, 18JM17693, 18JM17696, 18KM21705, 19BM02834, 19BM02839, 19CM19591, 19DM28090, 19DM28098, 19DM28099, 19EM19663, 19EM19670, 18GM06573, 18GM06581, 18DM18282, 18DM18284, 18EM17244, 18EM17247, 18HM16458, 18LM11838, 19BM02835, 19BM02836, 19BM02838, 19DM28092, 19DM28094, 19DM28096, 19DM28102, 19EM19669, 19CM19858, 19AM05185, 18GM20396, 18GM20401, 18GM20402, 18GM21563, 18KM09552, 19AM05177, 19AM05188, 18GM20401A, 18GM20402A  71338666, Lot Numbers: 18HM03836, 18JM15041, 18JM12829, 18JM12830, 19BM10927, 19BM10928, 19BM10931, 19BM10932, 18HM00329, 18HM00347, 18HM00347A, 18HM03834, 18HM03836, 18JM15041, 18HM03514, 18KM25967, 18LM15102, 19BM10935, 19BM10930, 18HM03834, 18KM09561, 18LM11826, 18KM06619, 18KM09560, 18KM09562, 18KM09563  71338667, Lot Numbers: 18HM12336, 18HM12361, 18HM12384, 18JM07422, 18KM09948, 19BM07657, 19BM07662, 19BM07666, 19DM15703, 19DM15708, 19DM15711, 18EM18375, 18GM06752, 18GM07679, 18HM00298, 18HM12366, 18JM07424, 18JM12833, 18JM12857, 19BM07656, 19DM15699, 19DM15702, 18HM12366A, 18KM17057, 18KM25976, 18GM06759, 18GM17009, 18HM12363, 18KM25971, 18JM00895, 18JM05154, 18KM25971A, 18KM14060  71338668, Lot Numbers: 18EM18377, 18EM18377A, 18HM13061, 18HM13079, 18HM13079A, 18HM13079B, 18HM22354, 19BM06037, 19BM06042, 19BM06043, 18HM00300, 18HM22357, 19EM19997

Root Cause

Process control

Action Taken

The firm notified its consignees by letter (delivered by Fed Ex) and email beginning on 04/22/2020. Consignees are asked to inspect inventory, locate and quarantine affected devices. Affected product will be returned to Smith & Nephew. Questions or concerns regarding this recall may be directed to: FieldActions@smith-nephew.com.