RecallDepth

LightSpeed Plus (4-slice MDAS LightSpeed 4.0 CT Scanner System - The LightSpeed 4.0 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. PET Discovery ST system - The GE Discovery ST system is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery ST is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery ST system can also be used as a stand-alone head and whole body multi-slice computed tomography (CT) diagnostic imaging system. LightSpeed Pro16 - The Lightspeed 5.0 CT Scanner System is indicated for head and whole body X-ray computed tomography applications. 16 Slice BGO PET/CT System - The GE Discovery XR and XA Systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery XR and XA are to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. These devices can also assist in radiotherapy planning. The Discovery XR and XA can also be used as stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging systems. 6 Slice BGO PET/CT System - The GE Discovery LS System is intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images. The Discovery LS is to be used by trained health care professionals for imaging the distribution of radiopharmaceuticals in the body for the assessment of metabolic (molecular) and physiologic functions. This can assist in the evaluation, diagnosis, staging, restaging, and follow up of lesions, disease and organ function such as (but not limited to) cancer, cardiovascular disease, and brain dysfunction. This device can also assist in radiotherapy planning. The Discovery LS can also be used as a stand-alone head and whole body multislice computed tomography (CT) diagnostic imaging system.

This recall has been terminated (originally issued February 10, 2016).

Company
Ge Medical Systems, Llc
Recall Initiated
February 10, 2016
Terminated
September 21, 2016
Recall Number
Z-1052-2016
Quantity
1
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

GE Healthcare has updated the Service Manuals for a number of CT and PET CT Scanners to update an incorrect torque value.

Distribution

Worldwide Distribution - US including all states in including PR except ND, VT & WY. Foreign Distribution: ALBANIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, BURKINA FASO, CAMERON, CANADA, CHILE, CHINA, COLOMBIA, DOMINICAN REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE,HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRA N, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMAHIRIYA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PERU, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TUNISIA, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA .

Lot / Code Info

Mfg Lot or Serial # System ID 00000002493PT5 0850450029

Root Cause

Labeling design

Action Taken

GE Healthcare sent an Urgent Medical Device Correction letter " GEHC Ref# 25467-ABT or 25467" dated February 12, 2016, to all affected consignees . The letter was addressed to Hospital Administrators / Risk Managers, Managers of Radiology/Cardiology, & Radiologist/Cardiologist. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Consignees were instructed to ensure that the organization that service their equipment uses the latest version of the Service Manual to maintain the safe and proper performance of their medical device. For questions call in the United States: 1-800-437-1171 or your local GE Healthcare Service Representative. For questions regarding this recall call 262-513-4122.

More recalls by Ge Medical Systems, Llc

View all recalls by this company →