NxStage System One S Cycler - Model no. NX1000-3. For home hemodialysis.

Company
Nxstage Medical, Inc.
Recall Initiated
October 29, 2015
Terminated
November 19, 2019
Recall Number
Z-0335-2016
Quantity
2,134 units
Firm Location
Lawrence, MA

Reason for Recall

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Distribution

Worldwide Distribution.

Lot / Code Info

Software version 4.9 and 4.10

Root Cause

Software design

Action Taken

NxStage issued a Field Safety Notice (FSN), dated October 29, 2015, to clinic and home patients with a NX1000-3 or NX1000-4 cycler. For patients performing short daily hemodialysis, the FSN included instructions for steps to be taken so that the cycler and treatments are not affected by the software error. Patients performing nocturnal hemodialysis were instructed to contact their Center for a replacement NX1000-1 Cycler. A software update will be released to correct the error. Customers should complete and return the reply form. Contact NxStage Customer Service at 1-866-NXSTAGE (1-866-697-8243) with questions or comments.