RecallDepth

Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

Company
Ferno-Washington Inc
Recall Initiated
August 3, 2015
Posted
October 1, 2015
Terminated
January 20, 2016
Recall Number
Z-2826-2015
Quantity
72 units
Firm Location
Wilmington, OH

Reason for Recall

Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

Distribution

Worldwide Distribution - US Nationwide in the states of: CO, IA, IL, IN, KY, MD, MN, MO, NC, NJ, NY, OH, OK, OR, TX, VA, WA & WY and the countries of CANADA, AUSTRALIA, UNITED KINGDOM & NEW ZEALAND.

Lot / Code Info

Serial #s: 15F001703, 15F001704, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001715, 15F001717, 15F001719, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001728, 15F001729, 15F001730, 15F001731, 15F001732, 15F001733, 15F001734, 15F001735, 15F001736, 15F001737, 15F001738, 15F001739, 15F001740, 15F001741, 15F001742, 15F001744, 15F001745, 15F001746, 15F001747, 15F001748, 15F001749, 15F001750, 15F001751, 15F001752, 15F001753, 15F001754, 15F001755, 15F001756, 15F001757, 15F001758, 15F001759, 15F001760, 15F001761, 15F001762, 15F001763, 15F001764, 15F001765, 15F001766, 15F001767, 15F001768, 15F001769 5F001770, 15F001771, 15F001772, 15F001773, 15F001774, 15F001775, 15F001776, 15F001777, 15F001778, 15F001779, 15F001780, 15F001781, 15F001782, 15F001783, 15F001784, 15F001785, 15F001786, 15F001787, 15F001788, 15F001789, 15F001790, 15F001791, 15F001793, 15F001794, 15F001795, 15F001796, 15F001797, 15F001798, 15F001804, 15F001805, 15F001806, 15F001807, 15F001808, 15F001809, 15F001810, 15F001811, 15F001812, 15F001813, 15F001814, 15F001815, 15F001816 & 15F001850.

Root Cause

Device Design

Action Taken

Ferno began calling their customers on 8/3/2015. The firm informed their customers of the following: To provide optimal Customer Experience and minimal disruption to your service, the iNX field action involves the following steps: " Exchange of your current iNX with a new iNX; " Participation in a brief training session; and " Inspection of your Ferno Model 175 antler and rail fastening system as requested by the Dept. of Transportation. As you await delivery of your new iNX, it is important you ensure proper function of the fastening system and iNX telescoping load frame.

More recalls by Ferno-Washington Inc