RecallDepth

FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.

Company
Ferno-Washington Inc
Recall Initiated
September 11, 2025
Posted
October 10, 2025
Recall Number
Z-0133-2026
Quantity
138 units
Firm Location
Wilmington, OH

Reason for Recall

The fastening post hardware could loosen allowing the fastening post to detach from the cot

Distribution

US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IS, IL, KY, LA, MN, MO, NC, ND, NE, NM, NY, OH, OK, PA, TN, TX, UT, VA, WI, WV.

Lot / Code Info

UDI-DI 00190790003210, Serial Numbers: 25G036127,25G036128, 25G036151, 25G036152, 25G036153, 25G036154, 25G036155, 25G036156, 25G036157, 25G036158, 25G036267, 25G036268, 25G036269, 25G036612, 25G036613, 25G036614, 25G036652, 25G036653, 25G036654, 25G036801, 25G036802, 25G036803, 25G036804, 25G036827, 25G036828, 25G036829, 25G036830, 25G036833, 25G036834, 25G036835, 25G036836, 25G036837, 25G036921, 25G036922, 25G036923, 25G036924, 25G036925, 25G036926, 25G036927, 25G036928, 25G036929, 25G036930, 25G036951, 25G036952, 25G036953, 25G036954, 25G036958, 25G036959, 25G036960, 25G036961, 25G036962, 25G036963, 25G036964, 25G037067, 25G037068, 25G037069, 25G037070, 25G037071, 25G037072, 25G037093, 25G037094, 25G037095, 25G037096, 25G037097, 25G037098, 25G037100, 25G037159, 25G037160, 25G037161, 25G037162, 25G037163, 25G037165, 25G037215, 25G037216, 25G037217, 25G037218, 25G037219, 25G037222, 25G037223, 25G037224, 25G037225, 25G037226, 25G037227, 25G037228, 25G037229, 25G037271, 25G037272, 25G037273, 25G037274, 25G037275, 25G037276, 25G037277, 25G037278, 25G037279, 25G037280, 25G037281, 25G037282, 25G037283, 25G037284, 25G037285, 25G037286, 25G037287, 25G037288, 25G037289, 25G037298, 25G037299, 25G037300, 25G037301, 25G037302, 25G037303, 25G037304, 25G037305, 25G037336, 25G037337, 25G037338, 25G037339, 25G037340, 25G037341, 25G037342, 25G037344, 25G037345, 25G037346, 25G037347, 25G037466, 25G037467, 25G037468, 25G037469, 25G037470, 25G037471, 25G037472, 25G037473, 25G037474, 25G037526, 25G037595, 25G037596, 25G037597, 25G037598, 25G037599, 25G037600, 25G037601, 25G037602, 25G037623, 25G037665, 25G037691, 25G037692, 25G037703, 25G037704, 25G037707, 25G037766, 25G037767, 25G037768, 25G037769, 25G037770, 25G037772, 25G037774, 25G037780, 25G037781, 25G037782, 25G037786, 25G038074, 25G038075, 25G038076, 25G038077, 25G038079, 25G038082, 25G038083, 25G038084, 25G038088, 25G038089, 25G038090, 25G038536, 25G038537, 25G038538, 25G038539, 25G038540, 25G038564, 25G038575, 25G038576, 25G038577, 25G038578, 25G038615, 25G038616, 25G038617, 25G038618, 25G038619, 25G038625, 25G038626, 25G038627, 25G038628, 25G038698, 25G038754, 25G038765, 25G038766, 25G038767, 25G038768, 25G038769

Root Cause

Device Design

Action Taken

FERNO issued an URGENT PRODUCT CORRECTION notice to its consignees on 09/11/2025 via USPS certified mail. The notice explained the issue, potential risk to user, and requested the consignee identify the affected unit, inspect it using the following criteria: Visual Inspection " Look for gaps between the fastener head and the surface. " Check for missing fasteners. " Look for shiny wear marks, rust streaks, or metal shavings around the fastener these can indicate movement. " Inspect for vibration marks (circular or elongated wear around bolt holes). Physical/Manual Check " Hand-check: Use your fingers or a small wrench to see if the fastener turns easily when it should not. " Shake or tap test: Lightly tap the fastener or surrounding component with a non-marring tool; rattling or movement suggests looseness. Functional Signs " Listen for unusual rattling or vibration during operation. " Check for misalignment of connected parts (hinges, brackets, covers, panels). " Watch for loosened thread-locking compounds if broken, the fastener may have shifted. IF A COT IS FOUND TO HAVE LOOSE OR MISSING HARDWARE, YOU MUST IMMEDIATELY REMOVE IT FROM SERVICE UNTIL THE REPAIR IS COMPLETE. Upon receipt of this Notice, please call Ferno Technical Support at 866-987-3776 or email tscoordinator@ferno.com to (1) acknowledge receipt of this notice, (2) advise if any of your Power X2 serial numbers are affected, (3) advise if and how many of your affected serial numbers contain loose and/or missing hardware on the fastening post, and (4) make arrangements for trained service technicians to correct your ambulance cot. In the event you sold or donated the identified product, please forward this notice to your consignee and advise Ferno of the consignee information so we may follow up. Malfunctions or quality problems experienced with the use of your POWER X2 Powered Ambulance Cot may be reported to Ferno at 866-987-3776 or tscoordinator@ferno.com