S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors.

Company
Aesculap Inc
Recall Initiated
July 7, 2015
Posted
August 3, 2015
Terminated
April 17, 2017
Recall Number
Z-2326-2015
Quantity
10
Firm Location
Center Valley, PA

Reason for Recall

A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Distribution

Distributed US (nationwide) and to Canada.

Lot / Code Info

product: FW103R, all Lots

Root Cause

Labeling design

Action Taken

Aesculap sent Field Safety Notification Letters, dated July 7, 2015, to end users, sales consultants, and distributors to advise them about the issue. Enhancements to the instructions for use were made available on the firm's website.