PHALINX (TM) HAMMERTOE HANDLE, SIZE: SMALL, STYLE: ANGLED, REF 45304012, CONTENTS 1 EACH, RX ONLY
- Company
- Wright Medical Technology, Inc.
- Recall Initiated
- April 16, 2015
- Posted
- May 18, 2015
- Terminated
- October 18, 2018
- Recall Number
- Z-1625-2015
- Quantity
- 121 units
- Firm Location
- Memphis, TN
Reason for Recall
The term DORSAL was laser etched on the incorrect side on some of the angled handles.
Distribution
Nationwide Distribution
Lot / Code Info
Lot 1344534
Root Cause
Process control
Action Taken
Wright sent an Urgent Medical Device Recall letter dated April 13, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory for the affected product, send it back to Wright Medical as soon as possible. Contact Customers Service at 901-867-4575 for return instructions and replacement inventory. Customers were instructed to complete the attached response immediately, confirming their receipt of this notice and the status of any product(s) on had, and return a copy by email to complaints@wmt.com or fax to 901-867-7401. Customers with questions should call 901-451-6144.