RecallDepth

stryker ORTHOLOC 3Di Fusion Plate MTP Right, REF 58A220RT, bone stabilization and fixation

Company
Wright Medical Technology, Inc.
Recall Initiated
April 10, 2023
Posted
April 26, 2023
Recall Number
Z-1466-2023
Quantity
41 units
Firm Location
Memphis, TN

Reason for Recall

Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.

Distribution

US Nationwide distribution.

Lot / Code Info

UDI/DI 00889797101011, Lot Number 1736809

Root Cause

Employee error

Action Taken

The firm issued an URGENT: MEDICAL DEVICE RECALL notice to its sales representatives and distributors on 04/10/2023 by email and followed with letters disseminated to the remaining consignees via tracked FedEx mailing on Wednesday, April 12. The notice explained the issue and the risk and requested the following actions be taken: Identify all affected products and isolate/quarantine to prevent accidental use. All parties that further distributed the product were directed to notify anyone to the recipients of the affected products. The firm is seeking return of the products.