GE Healthcare Centricity Perinatal (formerly QS) System - PFILS (Patient File Server) software; automatic patient data management providing clinical information at the bedside in Labor & Delivery, Mother-Baby and the Neonatal Intensive Care Unit.
This recall has been terminated (originally issued July 30, 2008).
- Recall Initiated
- June 30, 2008
- Posted
- July 30, 2008
- Terminated
- September 6, 2010
- Recall Number
- Z-2037-2008
- Quantity
- 1,031 units
- Firm Location
- Barrington, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Patient information mix-up: When the PFILS application, configured with a backup server has experienced a network interruption, PFILS verify could result in the Centricity Perinatal application recording non-identifiable patient information to the incorrect patient file.
Distribution
Worldwide: USA, Puerto Rico, Canada, and China.
Lot / Code Info
All CPN versions up to 6.70.6 P02 with PFILS versions below 2.0
Root Cause
Software design
Action Taken
GE Healthcare notified their Centricity Perinatal customers via letter dated 7/20/07 of the problem with the PFILS application that could result in the Centricity Perinatal application recording non-identifiable patient information to the incorrect patient file. The letter provided the users with a workaround and provided them with a Frequently Asked Questions document which listed the availability of a software patch to fix the problem. A follow-up letter dated 6/27/08 was sent on 6/30-7/2/08 to those accounts who had not requested a PFILS Hotfix CD-ROM or had not had their software recently upgraded. The letter a copy of the 7/20/07 letter, the most recent FAQ, the PFILS Hotfix CD (PFIL67_66) and Installation Procedure for the PFILS Patch instructions for installation of the software patch. The accounts were advised that they can request an analysis of existing patient records to determine whether they have been affected by the PFILS issue, by submitting a non-critical Technical Support Request on the GE Centricity Perinatal Web site: http://www.gehealthcare.com/usen/security/cis/stupgrades.html, indicating "PFILS ANALYSIS REQUEST" in the Description of Issue field. Any questions were directed to Technical Support at 800-433-2009.