RecallDepth

Centricity Enterprise Archive (EA) versions 3.0 to 3.0.7 software; GE Healthcare Integrated IT Solutions, Barrington, IL 60010. The software delivers a scalable and flexible DICOM storage solution for Cardiology and Radiology images.

This recall has been terminated (originally issued September 24, 2009).

Company
Ge Healthcare Integrated It Solutions
Recall Initiated
August 14, 2009
Posted
September 24, 2009
Terminated
February 25, 2011
Recall Number
Z-2316-2009
Quantity
212 units
Firm Location
Barrington, IL
Official Source
View on FDA website ↗

Reason for Recall

Software error: There is potential safety issue with Centricity Enterprise Archive (EA) 3.0.x software where study split operations are not correctly replicated to a secondary 'shadow' archive.

Distribution

Worldwide Distribution -- United States, Belgium, Canada, Cayman Islands, China, Denmark, France, Germany, Great Britain, Hong Kong, India, Israel, Italy, Korea, Malaysia, Malta, Mexico, New Zealand, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan and Turkey.

Lot / Code Info

Software versions 3.0, 3.0.1, 3.0.2, 3.0.3, 3.0.4, 3.0.5, 3.0.6 and 3.0.7

Root Cause

Software design

Action Taken

GE Healthcare Integrated IT Solutions sent Urgent Medical Device Correction letters dated August 7, 2009 to the attention to the Director of Radiology, the Hospital Administrator, the Head of the Radiology Department and the PACS Administrator. The letters advised users of the potential safety issue and provided safety instructions to follow until the software is updated. Users were asked to complete the enclosed Customer Reply form and fax the completed form to 1-847-939-1479. Any questions should be directed to the Remote Online Center or the GE Customer Care Center at 1-800-437-1171.

More recalls by Ge Healthcare Integrated It Solutions

View all recalls by this company →